Bio-Rad Model 583 Gel Drying Supports User Manual

Page 18

Advertising
background image

15

Maintenance and Troubleshooting

Table 6 (continued)

Troubleshooting Guide

Indication

Problem

Wet spot appears on

The vacuum pump was

the gel at the location turned off before the
of the vacuum outlet

vacuum seal was broken,

in the dryer base.

causing liquid to be pulled
back into the gel dryer
from the vacuum line.

Front panel display

The microprocessor is

flashes the message

unable to control the

HOT and the

heating element and the

alarm sounds.

temperature has
reached 100° C.

Front panel display

The microprocessor is

flashes the message

unable to keep the

rEg.

temperature stable at
the set value.

Action

Open the Gel Dryer and lift the transparent Sealing
Gasket prior to turning off the vacuum pump.

Press any key to clear the display. The dryer will
attempt to control temperature until the run is com-
pleted. An internal thermostat protects the dryer by
turning the heater off if it reaches 115° C.

Turn off the gel dryer power, and then turn it back on.
If the problem persists, contact your local Bio-Rad
office for repair.

Normal operation allows the dryer 20 minutes to
reach set temperature. If the dryer does not reach
the set value, or if it deviates from the set value by
more than 2° C during the run, it will register a regu-
lation error rEg.

Press any key to clear the display. The dryer will
attempt to control temperature until the run is com-
pleted.

Turn off the gel dryer power, and then turn it back on.
If the problem persists, contact your local Bio-Rad
office for repair.