Bio-Rad Sub-Cell® Model 192 Cell User Manual

Page 20

Advertising
background image

Do not use abrasive or highly alkaline cleaners on Sub-Cell parts. Do not expose Sub-Cell

parts to temperatures >60 °C. Do not sterilize Sub-Cell parts by autoclaving or dry heat.

4.3 Maintenance Schedule

Item

Look For

Frequency

Action

All Parts

Dried salts, agarose,

Each Use

Clean parts as

grease, and dirt

described in Section 4.1

Electrical cables

Breaks or fraying

Each Use

Replace Cables

Trays

Chips or cracks

Each Use

Replace Tray

Electrode Wires

Breaks

Each Use

See Section 4.4
(Electrode Cassette
Replacement)

Cable Connections

Looseness

Weekly

Replace Banana Jacks

(Banana Jacks

or Banana Plug Holders

and Plugs)

Base

Crazing, cracks,

Monthly

Replace Base

or leaks

Recirculation

Looseness or cracks

Each Use

Tighten or Replace

Tubing

4.4 Electrode Replacement

The Sub-Cell systems allow easy, hassle-free replacement of broken electrode wires by simply

removing the banana plug/electrode wire assembly and ordering a new assembly from Bio-Rad
(Figure 4.1). See Ordering Information (Section 6.2) for catalog numbers and part descriptions.

Instructions

1. Remove the thumb screw from the side wall of the base to release the banana plug/electrode

wire assembly. Do not discard this thumb screw (keep this screw with the base).

2. Remove the broken wire assembly from the base and discard the broken assembly.

3. Insert the new electrode assembly ensuring the electrode wire guard is properly seated into

the electrode wire guard slot in the bottom of the base.

4. Replace and tighten the thumb screw to secure the assembly in the base.

Figure 4.1 Replacement of banana plug/electrode wire assembly.

17

Advertising
This manual is related to the following products: