4.5 RNase Decontamination

Sub-Cell parts can be cleaned with a mild detergent and treated for 10 minutes with 3%

hydrogen peroxide (H

2

O

2

) and then rinsed with 0.1% DEPC- (diethyl pyrocarbonate) treated

distilled water to eliminate RNases prior to using the Sub-Cell systems for RNA gels

1-2

. Do

not soak Sub-Cell parts in DEPC water. Consult references

1-2

for other suggestions regarding

the use of DEPC in RNase decontamination.

Caution: DEPC is a suspected carcinogen. Always wear gloves, eye glasses and a
laboratory coat. Use caution when handling DEPC-containing solutions. Consult the
DEPC MSDS for more information.

Do not attempt to RNase decontaminate Sub-Cell parts using extreme dry heat.

Note: Several commercial products are also available for eliminating RNase contamination.
RNaseZAP

(Ambion) or RNase AWAY

(Molecular Bio-Products) are safe, simple and

effective methods that if used properly do not craze or fog the Sub-Cell parts. See
manufacturer instructions for proper use.