Bio-Rad Bio-Plex Pro Human Acute Phase Reagent and Diluent Kits User Manual

Page 24

Advertising
background image

20

1. Select Calibrate

and confirm that the default values for CAL1

and CAL2 are the same as the values on the Bio-Plex calibration
bead labels. Use the Bio-Plex low RP1 target value.

2. Select OK and follow the instructions for CAL1 and CAL 2 calibration.

Prepare Protocol
Lot-specific assay panel protocols are available for download at
www.bio-rad.com/bio-plex/standards
If the desired protocol is not available for download, create a new protocol:

1. Open a new protocol by selecting File, then New from the main

menu. Locate the steps at the left of the protocol menu.

2. Select Step 1 (Describe Protocol) and enter information about the assay.

3. Select Step 2 (Select Analytes) and choose the appropriate acute phase

panel, from the assay pull-down menu. If the panel is not listed, create
a new panel and select the “Bio-Plex Pro” assay type from the assay
pull-down menu. See the Bio-Plex Manager software instruction
manual for details.

NOTE: It is important to select the appropriate assay type in the assay
pull-down menu to ensure that data is acquired using the correct bead
map and gate settings.

4. Choose the analytes to be selected for the assay from the available

list of targets. Note the targets will already be entered in the selected
view when using a downloaded protocol. In future runs, the selected
panel can be copied into the “available” view.

5. Select Step 3 (Format Plate) and click on the Plate Formatting tab.

Select the number of replicates, the autofill direction, and click on
the

icon. Drag the cursor over all the wells that contain standards.

Repeat by replacing the

with

,

, and

to identify all the

wells that contain blanks, controls, and samples, respectively.