Bio-Plex Manager Software Version 4.1 and 4.1.1

Prepare System
1. Empty the waste bottle and make sure that there is sufficient sheath

fluid in the bottle or cube before proceeding.

2. Turn on the reader, XY platform and HTF (if included). Allow the

system to warm up for 30 min.

3. Select Start up

and follow the instructions to prepare the reader

to acquire data. If the system is idle for 4 hr without aquiring data, the

lasers will automatically turn off. To reset the 4-hour countdown, select

Warm up

and wait for the optics to reach operational temperature.

Calibrate With Low RP1 Target Value
Calibrate using Bio-Plex calibration beads and the

low RP1 target on the

CAL2 bottle. This software version requires the user to calibrate either on

the low or high RP1 target value. Daily calibration is recommended before

acquiring data.

1. Select Calibrate

and confirm that the default values for CAL1 and

CAL2 are the same as the values on the Bio-Plex calibration bead

labels. Use the Bio-Plex

low RP1 target value.

2. Select OK and follow the instructions for CAL1 and CAL2 calibration.

Prepare Protocol
1. Open a new protocol by selecting File, then New from the main

menu. Locate the steps at the left of the protocol menu.

NOTE: To minimize protocol setup, previously prepared lot-specific

Bio-Plex Pro acute phase assay protocols in BPM 4.0 are available

for download at

www.biorad.com/bio-plex/standards. The user

will need to select the 25 region bead map in Step 7 (Run Protocol).

2. Select Step 1 (Describe Protocol) and enter information about the

assay.