7. Select Step 5 (Enter Controls Info) to enter controls information. This

is where the concentration of the user-specified controls is entered

into the protocol.

a)

If necessary, deselect the box for same concentration values for

all analytes.

b)

Select an analyte from the pull-down cell.

c)

Enter the description and dilution information for each user-

specified control.

d)

Repeat steps 7b and 7c for each additional analyte in the assay.

8. Select Step 6 (Enter Sample Info) and enter sample information.

Remember to enter the appropriate dilution factor information.

9. Select Step 7 (Run Protocol). Then:

a)

Choose

100 beads per region.

b)

In Advanced Settings, set the bead map to

25 region.

c)

In Advanced Settings, set the sample size to

50 µl.

d)

In Advanced Settings, confirm that the DD gate is set to

5,000

(low) and

32,000 (high).

e)

Save the protocol (.pbx) file.