Bio-Rad Model EM-1 Econo™ UV Monitor User Manual

Page 16

Advertising
background image

13

Problem

Possible cause

Solution

Lamp will not

Optics module not

Plug optics module

light,or lamp LED plugged in

into the

socket

on continues blinking

rear of UV moni-

tor.

Lamp burned out

Replace lamp (see

Appendix B).

Chart recorder

Chart recorder is

Plug chart recorder

not responding

plugged into AUX

into the

socket.

socket

Full scale of recorder

Set recorder range to

set too high

1 V.

Recorder has bottomed

Readjust pen position.

out

Power adaptor not

Check that recorder

plugged in

power adaptor is

plugged in properly.

No power

Faulty power

Check power cable

connection

connection.

Blown fuse

Check fuses.

M7318160G.qxd 05/08/2003 2:05 PM Page 13