Bio-Rad Model EM-1 Econo™ UV Monitor User Manual

Page 20

Advertising
background image

17

Appendix C
Rear Panel Connections

Both the signal

output and the AUX sockets are 8-pin circular

mini-DINs. The following information is provided for those wishing to
interface the Model EM-1 Econo UV Monitor with other equipment
such as integrators and non Bio-Rad fraction collectors. System Cable 7
may be used as a breakout cable, with an 8-pin mini-DIN at one end and
loose wires at the other.

Do not attempt to use the

socket for anything other than the

Model EM-1 Econo UV Monitor optics unit.

All digital inputs use TTL circuitry and are active LOW.

Pin #

Function

Notes

1

Connects to

Paper stop

AUX Pin 1

2

No Contact

3

No Contact

4

Integrator (+)

0 to 1 V = 0 to 2 AU

5

Chart Recorder (+)

0 to 1 V = Full scale

6

Connects to

Pen down

AUX Pin 6

7

Ground

8

No Contact

Pin #

Function

Notes

1

Connects to

Pin 1

2

Auto-Zero

3

Large Mark

Marks chart recorder

signal only

4

Integrator (+)

0 to 1 V = 0 to 2 AU

5

No Contact

6

Connects to

Pin 6

7

Ground

8

Small Mark

Marks chart recorder

signal only

pin

configuration

pin

configuration

M7318160G.qxd 05/08/2003 2:06 PM Page 17