Bio-Rad Model EM-1 Econo™ UV Monitor User Manual

Page 18

Advertising
background image

15

Appendix B
UV Lamp Replacement

The coated low pressure mercury lamp has an expected lifetime of

at least 2,000 hours. To replace a lamp, proceed as follows, referring to
Figure B-1:

1. Unplug the optics module from the rear panel of the control unit.

Warning: Failure to unplug the optics module from the control unit
may expose the user to high voltage and UV radiation.

2. If the UV lamp has been on, allow the optics module to cool for 10

minutes before continuing.

3. Remove the filter tray (see Section 5.2) and the flow cell (see

Appendix A).

4. Remove the screws holding the optics module together and remove

the bottom half of the case.

5. Pull the old UV lamp out of its holder and unplug it from its

connector.

6. Plug the new UV lamp into its connector and insert it into the lamp

holder. Never handle the quartz surface of the UV lamp. Grease and
fingerprints will destroy the lamp.

7. Reassemble the optics module (do not over-tighten the screws).

!

M7318160G.qxd 05/08/2003 2:05 PM Page 15