18

Appendix D
Technical Specifications

Wave length

254 nm and 280 nm

Operating mode

AU

Lamp

Low pressure mercury lamp with phosphor

screen

Filters

Interference type

280 nm

254 nm

Operating temperature +4° to 40° C
Baseline offset

± 0.25 AU Auto-Zero

Time constant

1.5 sec to 90% of full response at all ranges

Sensitivity ranges

2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01 AUFS

Output signal

0-1 V, impedance 150 ohms

Power requirements

90-132 VAC (120 V at delivery)

180-265 VAC (220 V at delivery)

Power consumption

15 watts

Dimensions

Base unit 14.55 x 18.64 x 20.24 cm

(W x H x D)

Optics unit 13.21 x 15.24 x 3.81 cm

(W x L x H)

Weight

Base unit 2.2 kg (4.8 lb)

Optical unit 311 g (11 oz)

Flow cell

Optical path 2 mm

Internal volume 80 µl

Illuminated volume 3 µl

Noise

1.0 x 10-4 AU max peak to peak (dry cell)

2 x 10-4 AU max peak to peak (flowing liq-

uid)
Linearity

3% to 2 AU at 254 nm

Long term drift

< 1 mAU/hr-deg C

Warm-up time

1 hour to meet drift specification

Flow sensitivity

<200 µAU/(ml/min)

Regulatory

CSA: C22.2 No.151-M1986

Compliance:

TÜV: EN61010-1: 1993
CE:

EN55011:1991 Class A
EN50082-2: 1995