Bio-Rad Bio-Scale™ Mini UNOsphere SUPrA™ Affinity Cartridges User Manual

Page 10

Advertising
background image

Table 2. Technical Description of UNOsphere
SUPrA Affinity Chromatography Media.

Composition

Highly crosslinked polyacrylamide polymer

Particle size range

53–61 µm

Ligand

Recombinant protein A

Coupling chemistry

Epoxy

Dynamic binding capacity*

150 cm/hr

30 ± 3 mg/ml

300 cm/hr

25 ± 2 mg/ml

450 cm/hr

20 ± 2 mg/ml

Chemical stability**

10 mM hydrochloric acid

6 M guanidine hydrochloride

0.1 M arginine (pH 2.8)

0.1 M citrate (pH 2.8)

0.1 M glycine (pH 2.8)

Working pH range

3–11

Cleaning-in-place (CIP)

6 M guanidine hydrochloride

10 mM hydrochloric acid

0.1 M sodium hydroxide

1 M acetic acid/20% ethanol

Recommended mobile phase

100–600 cm/hr

velocity range

Temperature stability

4–40°C

Delivery conditions

50% slurry in 20% ethanol

Storage conditions

4–8°C

*

Minimum 20 mg/ml at 300 cm/hr; 10% breakthrough capacity determined

with 1.0 mg/ml polyclonal human IgG in 1.1 x 10 cm column.

** No significant change in chromatographic performance after 24-hour storage

at room temperature.

5

10014997A:10005454A.qxd

9/23/2008

3:16 PM

Page 5

Advertising