3. Connect the inlet of the cartridge to the male

Luer fitting on the MV-6 sample inject valve. (See
Figure 2.) If not using the MV-6 sample inject
valve, connect a barb to male Luer fitting on the
1.6 mm ID tubing, then connect to the top of the
female Luer on the Bio-Scale mini cartridge. For
optimum performance, a cartridge should be
mounted vertically with the arrow on the
cartridge pointing downward.

4. Connect the cartridge outlet to the 1.6 mm ID

tubing leading to the BioLogic LP optics module
or Econo UV monitor. It is recommended to use
the shortest length (approximately 10 cm) of
1.6 mm ID tubing. Connect a barb to female
Luer fitting to the 1.6 mm ID tubing, then
connect to the bottom of the male Luer on the
Bio-Scale Mini cartridge.