Section 6
Regenerating the Column
A 0.4 M phosphate buffer is usually sufficient to
remove adsorbed materials. If a contaminated sample has
been run, it is advisable to remove the top layer of the
hydroxyapatite bed and wash the remainder of the bed
with one bed volume of 1 M sodium chloride followed
by four column volumes of starting buffer.