Integra LifeSciences Laparoscopic Instrument Care and Cleaning User Manual

Page 5

Advertising
background image

5

Thoroughly inspect instruments for effective cleaning (free

from any visible dirt and residue). Also inspect the

instruments for any damaged, worn, corroded, pitted/porous

conditions and ensure proper function. Contact Miltex if

damage is observed and if repair is necessary. Apply

non-silicone instrument lubricant (such as Miltex REF 3-700

Instrument Spray Lube) to the joints and jaws of the instrument

to ensure easy movement. Caution: Do not oil electrical

connections on instruments.

Ultrasonic Cleaners:

Ultrasonic Cleaners can also be used for cleaning and removal

of contamination and debris. Follow the Ultrasonic Cleaner

manufacturer’s recommendations for cleaning agents, fill

levels, temperature, processing time, etc. Make sure that

functional parts do not touch other instruments and all jointed

instruments are processed in an open position. We

recommend pre-flushing instruments via the equipped

LUER-LOCK connection which should be opened and force flushed

using the appropriate equipment prior to immersion in the

ultrasonic bath. After cleaning and disinfection, the

instruments should be thoroughly rinsed, both on the exterior

and interior sheath, with distilled water. Immediately dry

instruments to prevent corrosion. Drying can be performed

manually, but compressed air pistol is recommended for more

delicate instruments and is more effective for overall drying.

Thoroughly inspect instruments for effective cleaning (free

from any visible dirt and residue). Also inspect the

instruments for any damaged, worn, corroded, pitted/porous

conditions and ensure proper function. Contact Miltex if

damage is observed and if repair is necessary. Apply

non-silicone instrument lubricant (such as Miltex REF 3-700

Instrument Spray Lube) to the joints and jaws of the instrument

to ensure easy movement. Caution: Do not oil electrical

connections on instruments.

Warning: Ultrasonic cleaning should never be used for

endoscopes and instruments with tungsten carbide inserts.

2. STERILIZATION

Steam sterilization should be carried out by steam under

pressure in the complete absence of air in an autoclave set at

134º C as follows:

• Temperature 134°C, Sterilizing time = 5 min.

Instrument jaws and ratchets must be in open position to avoid

damage during the sterilization heating and cooling process.

Always follow the autoclave manufacturer instructions and

recommendations. For autoclaving, place the instruments in a

suitable container or sterilization tray. Cutting edges and tips

of instruments should always be protected with a soft cloth.

After autoclaving allow all components cool down to room

temperature. Sudden changes in temperature may cause

damage to the instruments.

3. GAS STERILIZATION

Place the pre-cleaned instrument in a transparent aseptic

package. It is recommended to cover tips and cutting edges of

instruments to avoid perforation of the packaging. Follow the

instruction manual of the manufacturer of the gas sterilization

equipment and observe domestic health care regulations.

Allow sufficient time for ventilation to remove toxic residues.

4. SERVICE

Contact the Miltex Repair department at toll free phone

number 1-866-854-8300. Miltex does not supply its

mechanical parts to any outside company, therefore only Miltex

or its subsidiaries are able to repair Miltex

®

brand laparoscopic

instruments with original parts and to guarantee the original

technical specification and safety of the product.

Additionally, this warranty is void for Miltex

®

instruments

and products serviced by any person or facility other than

the Miltex Instrument Repair Services.

5. WARRANTY

Miltex

®

laparoscopic surgical hand instruments have a 1-year

limited warranty against defects in workmanship and material

which excludes wear to blades and jaws. Any Miltex

laparoscopic instruments that prove defective in

workmanship or material will either be repaired or replaced

at our discretion, without charge. Our liability under this

warranty shall be limited to repair or replacement of defective

merchandise. This warranty is a limited warranty and is the

only warranty made by Miltex. Miltex makes no other warranty,

expressed or implied, including any warranty of merchantability

and fitness for a particular purpose.

INSTRUCTIONS FOR CARE & CLEANING OF

MILTEX

®

ENDOSCOPES AND LAPAROSCOPIC

SURGICAL HAND INSTRUMENTS

Instructions for Use/English

Advertising