Integra LifeSciences Care and Handling of Padgett Instruments User Manual

Indications for use, Contraindication, Warning

Advertising
background image

INDICATIONS FOR USE

Padgett surgical instruments are designed to perform a specific

function, such as cutting, grasping, clamping, dissecting, probing,

retracting, draining, aspirating, suturing, or ligating. For use by,

or as directed by, a surgeon. Instruments should be used only

for the purpose for which they are designed. The proper surgical

technique for the use of instruments is the responsibility of the

surgeon.

CONTRAINDICATION

Instruments should not be used for anything other than their

intended use.

WARNING

Consult individual national infection control/prevention protocols

for specific guidance regarding processing medical devices with

suspected exposure to Creutzfeldt-Jakob Disease (CJD).

CAUTION

After cleaning, especially ultrasonic cleaning, check screws on

instruments because the vibration from the ultrasonic cleaning may

cause them to loosen or fall out.
Padgett surgical instruments are supplied non-sterile, unless

otherwise noted, and must be cleaned, lubricated and sterilized

prior to use according to hospital protocol and the procedures

outlined in this document. Failure to follow these procedures will

invalidate the instrument’s warranty and can cause the instrument

to fail.
Inappropriate use of instruments will lead to damage that is usually

not repairable; for example, a hemostat that is used to clamp

tubing can become misaligned and quickly break.

INSPECTION OF ALL INSTRUMENTS

All instruments are carefully inspected before shipment. Because

damage may occur during transit, the instruments should be

thoroughly inspected upon receipt. All instruments must be inspected

prior to use.
Handling and Operating Instruments: Instruments should be handled

and operated by personnel completely familiar with their use,

assembly and disassembly. Before a new instrument is used

and prior to each surgical procedure, the instrument must be

decontaminated, lubricated and sterilized as described below.

Handle the instrument with care. The instrument must be inspected

to assure proper functioning prior to each use with particular

attention paid to the condition of all moving parts, tips, box locks,

ratchets and cutting edges. Each instrument with a screw must

be inspected before and after use to ensure that the screws do

not move when operating the instrument. Screws can loosen and

back out of an instrument as a result of normal operation and/or

the vibration during ultrasonic cleaning. Failure to make a complete

inspection to assure the proper operation and function of the

instrument may result in unsatisfactory performance, perhaps

because a part is missing. Do not use if the instrument does not

appear to be functioning properly. Use of an instrument for a task

other than that for which it is intended could result in a damaged

or broken instrument, or one which provides an unsatisfactory

performance. In order to insure warranties and guarantees,

instruments in need of repair should be sent to Integra.

DECONTAMINATION AND STERILIZATION PROCEDURES

As with any decontamination procedure, personnel should follow

accepted guidelines for hand washing, the use of protective attire,

etc. as recommended by A.A.M.I. Standards and Recommended

Practice, “Safe Handling and Biological Decontamination of

Medical Devices in Health Care Facilities and in Non-Clinical

Settings”, ANSI/AAMI ST35:2003.
Decontamination is a two step process:
1) Thorough cleaning and rinsing.
2) Sterilization or disinfection.

A. MANUAL DECONTAMINATION

PRECLEANING: Remove gross debris from surgical instruments with

a lap sponge and sterile water routinely during the procedure to

prevent drying on of blood and body fluids, etc. It is important

to rinse instruments that have been exposed to blood and saline

solution before these substances dry. Blood and body fluids as

well as saline solutions are highly corrosive. In addition, blood can

produce a stain that is difficult to remove.
CLEANING: To prevent the formation of biofilm, cleaning should

occur as soon as possible after instrumentation is used. Biofilm

is an accumulation of a biomass of bacteria and extracellular

material that tightly adheres itself to the surface of the instruments.

It cannot be easily removed, and protects microorganisms from

being easily removed by ordinary cleaning/decontamination

methods used in hospitals. It is particularly problematic in lumened

medical devices.
Step 1. Maintain moisture: Immediately after the surgical procedure,

place the instruments in an instrument tray/container and cover

with a towel moistened with sterile distilled water. Foam, spray

or gel products, specifically intended for use with surgical

instruments, are available to keep the soil moist. Transport tray of

soiled instruments in an impervious plastic bag or container with a

tight lid to the decontamination environment (keep the outside of

the containment clean).
Step 2. Enzymatic Soak: Immerse fully opened and/or disassembled

instruments in an enzymatic solution, specific for use with

surgical instruments. Prepare the solution and use per enzyme

manufacturer’s recommendations, paying special attention to

instructions for correct dilution, temperature and soak time.

Flush air from lumens and fill them with enzymatic solution for full

contact with this inner surface during the soak time.
Step 3. Rinse: Remove from enzymatic soak after the time period

recommended by the enzymatic manufacturer and rinse thoroughly

with tap water. Flush lumens until rinse water runs clear.
Step 4. Cleaning Instruments: Choose a cleaning solution

appropriate for surgical instruments and follow the manufacturer’s

instructions for use. The use of neutral pH detergents is vital to

the maintenance of surgical instruments. Contact with acidic or

alkaline solution will remove the instruments’ protective barrier

of chromium oxide, often leading to corrosion, pitting, and

breakage. You may find that depending on the type of soil, a

detergent that is a little more or less acid or alkaline may be more

appropriate. The ideal cleaning agent is nonabrasive, low-foaming

and free-rinsing. Using a small clean hand-held brush, remove

soil from all surfaces of the instrument while fully immersed in

the solution. During manual cleaning, never use steel wool, wire

brushes, scalpel blades or highly abrasive detergent or cleansers

to remove soil from surgical instruments. These will damage

the instruments’ protective surface and lead to corrosion. Use a

clean soft bristled brush to clean instruments with an accessible

channel. Remove the soil from the ratchets, jaws, tips, box locks,

and/or hinge mechanism. The box lock and hinge portion of an

instrument must be thoroughly cleaned after each use. A build-

up of soil, debris, lubricants, etc. in these areas, will make it

difficult to use the instrument and eventually irreparably damage

it. Vigorously flush channels with the cleaning solution. Deionized

water is recommended and preferred because it is free of the many

compounds which exist in ordinary tap water. These substances,

alone, cause stains and when tap water is combined with some

detergents it will form insoluble deposits on the instruments.

Manual cleaning should remove all visible residue. It is essential

to keep the box locks and hinges open during any manual or

automated cleaning process.
Step 5. Rinse: Thoroughly rinse instruments by immersing in tap

water and wiping with a clean, soft cloth. Flush lumens until water

runs clear.
Step 6. Ultrasonic Cleaning and Rinsing: Follow the recommendations

of the ultrasonic manufacturer regarding cycle times, detergents,

proper placement of the instrument tray, and conditioning

(“degassing”) of the cleaning solution, etc. Use an ultrasonic

cleaner to remove soil from hard to reach surfaces such as

grooves, crevices, lumens, instruments with moving parts,

etc., after gross soil has been removed. Open or disassemble

instruments as appropriate. Place instruments in a mesh bottom

stainless steel instrument tray. Place the tray into the ultrasonic

cleaner. Flush air out of lumens and fill them with the ultrasonic

cleaning solution (under the solution level in the chamber) for

effective removal of soil from that inner surface by the ultrasonic

activity.
Step 7. FINAL RINSE should be with “treated water”. Softened

or deionized water should be used for the final rinse to better

remove detergents etc. Softening water removes calcium and

magnesium ions that cause water to be hard. Iron ions may also be

removed by this treatment. Deionization removes ionized salts and

particles from the water. Excessively hard water can spot or stain

instruments and excessive chlorine in water can cause pitting of

the instrument. Deionized water is preferred for the final rinse.
Step 8. Decontaminate Clean Instruments: Once instruments have

been cleaned they must be rendered safe for handling, inspection

and assembly. They may be steam sterilized without a wrapper or

disinfected following the instructions from the instrument, sterilizer

and disinfectant manufacturers.
Step 9. Visual Inspection and Instrument Set Assembly: Visually inspect

the instrument for cleanliness and to ensure all parts are in proper

working order, as the set is assembled. Inspection is a vital part of

proper care and maintenance. Instruments in need of repair will not

perform accurately in surgery and breakage is likely to occur. DO

NOT USE damaged instruments. Worn ratchets, loose box locks

Advertising