Setup screen – Hudson Robotics Bar Code Print and Apply User Manual
Page 15

Print and Apply User Manual
Hudson Robotics, Inc.
13
B. Setup Screen
The unit requires setup in 3 steps to ensure successful use:
Label Retrieval Setup: Calibration of how the unit picks up labels, which
includes calibration of the vacuum and airblast, and position of the label
head when retrieving a label.
Apply Nest Setup: Calibration of the rotational nest on the applicator, as
well as the position of the applicator head in relation to the nest when
applying labels.
Apply Profile Setup: Calibration of plate-specific application profiles, to
ensure proper placement of labels during operation. (This is covered in
Chapter 4: Operation)
Setup cannot be run in simulate mode. All functions in setup require the print and apply
unit to be active and connected to the computer.