3.15.1 – Micro10x
Selecting the Micro10x action for this step will change this step’s display
to appear as follows:
This option allows a SOLOPlus to perform rapid dispensing from an
attached Micro10x. The user can select to run a dispense or prime
program, or to empty the fluid back into its source. Dispense and prime
programs are configured ahead of time in the Micro10x itself. Consult the
Micro10x manual for more information on creating and editing dispense
and prime programs.
3.15.2 – Ten WayValve
Selecting the Ten Way Valve action for this step will change this step’s
display to appear as follows:
This option allows a SOLOPlus with an attached Ten Way Valve input for
the Micro10x to modify the fluid being dispensed by the Micro10x.
Execution of the step results in a change in the input valve port on the Ten
Way Valve. It is recommended to re-prime the Micro10x after switching
fluids. Optionally, the user may want to empty the Micro10x before
changing fluid sources to conserve liquids.