Input the external trigger signal

The external trigger signal is input from the digital I/O connector.

(See 5-1 ‘Connector Pin Arrangement)

Connector pin arrangement

Pin

No.

Monitor Board

Signal Name

I/O Signal

Description Logic Remarks

B5

10 Input

External

Trigger

Signal

High: ON

See Circuit below.

Connection Specifications

3.3V TTL Input

Operation procedure

(Raster Scan Parameter Settings)

Set Parameter ID =26

Raster Scan Time

Setting

Set Parameter ID =27

Raster Scan Duty Ratio

Setting

Set Parameter ID =28

Raster Scan Angle

Setting

(Raster Scan Operation)

Send Command ID=8

Data=6 Movement Start

(Movement to the initial position

of

a

raster

scan)

Send Command ID=23 Data=8 (One way scan)

Data=9 (One coming and going)

(Reference: Data = 3 in case of continuous operation)

Input the external trigger Signal

Send Command ID=23

Data=0 Scan Stop

Notes

The external trigger input during raster scan move is ignored.

From the external trigger signal input, until the time actual operation starts, there can be a

fluctuation of up to a maximum of 10 usec.

FPGA