Tuning – Canon GM-1000 User Manual

Page 76

Advertising
background image

CANON Digital Galvano Scanner System GM-1000 Series
Users Manual 1.20

76

7.

Tuning

7.1. Tuning

At the time of shipping, with the combination of the galvano motor and controller (GC-201), appropriate

servo tuning has been completed.

The following tools have been prepared to check malfunctions during use, and handling trouble such

as mirror damage.

Control software

It is possible to check servo tuning related parameters and make changes from the control

software’s Servo Setting tab.

Note: Changes are not usually made to the parameters. Settings with inappropriate parameter

values can cause abnormal operations such as oscillations.

Frequency characteristic (FFT) measurement

It is possible to check the servo tuning status.

(See 7.2 Frequency Characteristic (FFT) Measurement for details)

Auto tuning

When adjusting servo tuning, in order to exchange a damaged mirror, an easy auto tuning

program is required.

However, use of this function is limited to the mirror shape provided by the setting file (definition

file).

(See 7.3 Easy Auto Tuning for details)

X and Y axis matching

This function automatically matches the step movement response waveform of Axis 1, and Axis 2.

(See 7.4 X, Y Matching for details)

Advertising
This manual is related to the following products: