Chapter 5 Performing ECG Tests
Welch Allyn CP 100 Electrocardiograph
Figure 28. Electrode Placement Locations — Augmented Pediatric
Electrodes
AHA
IEC
Locations
A
V1
red
C1
red
Fourth intercostal space at right sternal border.
B
V2
yellow
C2
yellow
Fourth intercostal space at left sternal border.
C
V4
blue
C4
brown
Fifth intercostal space at left of midclavicular line.
D
V7
orange
C7
black
Posterior axillary line at the same horizontal level as V4.
E
V6
purple
C6
purple
Mid-axillary line on same horizontal level as V4.
F
LA
black
L
yellow
Just above left wrist on inside of arm.
G
LL
red
F
green
Just above left ankle.
H
RL
green
N
black
Just above right ankle.
I
RA
white
R
red
Just above right wrist on inside of arm.
J
V3R
green
C3R
green
Mirror position to standard V3 placement with respect to the mid-
sternal line.
I
A
H
B
J
C
E
D
F
G
To use the augmented pediatric lead set, you
must enable it in the settings. See