Replacing the battery (dc) fuse, Replacing the battery, Dc) fuse – Welch Allyn CP 100 12-Lead Resting Electrocardiograph - User Manual User Manual
Page 67

Directions for Use
Chapter 6 Maintaining the Electrocardiograph
63
9.
Replace the battery door and screws. Turn the electrocardiograph rightside-up.
It turns on automatically and displays a prompt to enter date and time.
10. Enter the date and time.
The electrocardiograph is ready to use.
11. Discard the old battery appropriately.
•
In the USA, call 1-800-SAV-LEAD for instructions on how to recycle it.
•
International users, contact your local authorities concerning recycling.
Replacing the Battery (DC) Fuse
If the battery (DC) fuse requires replacing, as described in
Step 5
on page 61, follow these
steps. For the fuse value, see
1.
Remove and discard the fuse. See
.
You may need to pull hard. Use pliers gently if necessary.
2.
Connect a new fuse. It goes in either way.
3.
Go to
Step 8
on page 62.
Figure 43. Battery Fuse Removed