Specifications, A - specifications, Fuses – Welch Allyn CP 100 12-Lead Resting Electrocardiograph - User Manual User Manual
Page 77

A
73
Specifications
Item
Specification
Dimensions
Weight
11.6 lb (5.3 kg)
Keyboard type
Elastomer keypad with complete alphanumeric keys
Paper type
8.25 x 11 inches (21 x 28 cm), Z-fold thermal paper, 200 sheets
Thermal printer (internal)
Computer-controlled dot array, 8 dots/mm
USB printer (external)
Inkjet or laser printer that supports PCL (printer control language)
Thermal paper speeds
10, 25, 50 mm/s
Gain settings:
Auto ECGs
Rhythm ECGs
5, 10, 20 mm/mV, AUTO
5, 10, 20 mm/mV
Report print formats:
Auto ECGs
Rhythm ECGs
Average cycles
3x4, 3x4+ 1R, 3x4 +3R, 6x2, 12x1, 6x2 50 mm/s, 6x2 Ext., 2x6 +1R, 6x2 +1R
3 leads at a time
3x4 50 mm/s + 3R, 6x2 50 mm/s + 6R
Frequency range
0.3 to 150 Hz
Digital sampling rate
> 1,000 samples/second/channel
Pacemaker detection
ANSI/AAMI EC11
Power requirement
Universal AC power supply 100–240 V ~, 50/60 Hz ~, 2.2 A maximum
Fuses
AC
DC (battery fuse)
Time-delay type, 2 amp 250 V rating, Littlefuse 0215002 or equivalent.
Fast-acting type, 10 amp 32 V rating, Bussman ATC-10 or equivalent.
Lead configurations
Standard, Cabrera
Rechargeable battery
Lead acid gel, 6 volt, 5 AH
Prints up to 100 continuous ECGs per charge
12-hour recharging
15.6 in
(39.7 cm)
6.2 in
(15.7 cm)
16.2 in
(41.1 cm)