Electromagnetic Emissions and Immunity Information

Electromagnetic Emissions

The Welch Allyn Holter Recorder is intended for use in the electromagnetic environment specified below. The customer or user of the Welch
Allyn Holter Recorder should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions
CISPR 11

Group 2

The Welch Allyn Holter Recorder must emit electromagnetic energy in order to perform
its intended function. Nearby electronic equipment may be affected.

RF emissions
CISPR 11

Class B

The Welch Allyn Holter Recorder is suitable for use in all establishments, including
domestic establishments. The recorder has no connection to the public low-voltage
power supply network.

Harmonic emissions
IEC 61000-3-2

Not applicable

Voltage fluctuations/
flicker emissions
IEC 61000-3-3

Not applicable

Electromagnetic Immunity

The Holter recorder is intended for use in the electromagnetic environment specified below. The customer or user of the Holter recorder should
assure that it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic discharge
(ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.

Electrical fast transient/
burst
IEC 61000-4-4

±2 kV for power supply
lines
±1 kV for input/output
lines

Not applicable.

Surge
IEC 61000-4-5

±1 kV differential mode
±2 kV common mode

Not applicable.

Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines.
IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle
40 % U

T

(60 % dip in U

T

)

for 5 cycles
70 % U

T

(30 % dip in U

T

)

for 25 cycles
<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Not applicable.

Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8

3 A/m

Not applicable.

Note: U

T is the a.c. mains voltage prior to application of the test level.