Directions for Use
59
Failed to print
Printer Off-line
Printer out of paper
Printer not connected
Confirm that printer is online.
Confirm printer has paper.
Confirm printer connected.
Using another application, test the printer.
Make sure proper printer is selected with application.
Failure during File transfer
Loss of network connection
User prompted to “Retry” or “Abort”
“Retry” will attempt to reestablish network connection and
continue to transfer file.
“Abort” will lose all unsaved changes and exit the Holter
Application.
Condition
Cause
Action
Table 23. Holter Recorder Acquisition Errors
Recorder LCD displays the error code and Yellow LED (HR100) flashes the error code,
pause for two seconds, and repeats. Remove the battery to repeat diagnostic tests.
Code
LCD and LED
Description
E1
Not used
E2 - Initialize
Configuration error – recorder SD card contains data or awaiting Bluetooth initialization
E3 - BATT
Low/weak battery
E4 - SD Card
Secure digital card not detected
E5 - Call Service
Internal data processing error
E6 - Call Service
Power on self test failed
E7 - Call Service
Wireless data link error
E8 - SD Card
Error writing file to secure digital card
E9 - SD Card
Error reading file from secure digital card