Module OEM Implementation Manual

Welch Allyn OEM Technologies

Confidential

Page 33

Host Command

Description

<Qxxy>

Function enable

Allows the host to enable or disable functions listed below.

<Qxxy>

where xx (ASCII) is the compensation and y (ASCII)
is ON/OFF status.

xx =

00h

O

2

compensation

01h

N

2

0 compensation

02h

desflurane compensation

08h

water vapor compensation

09h

BTPS compensation

0A

Baseline compensation

7F

protect mode

y =

00 binary

OFF

01 binary

ON

3Fh

report current ON/OFF status

Example:

<Q011>

compensate the CO

2

waveform data for N

2

O

<Q7F0>

turn off protected mode (to allow certain commands)

<F20yy>

Change pump flow rate

Allows the host to specify the pump flow rate in milliliters per minute.

<F20yy>

where yy (8 bit) is the current pump flow rate.

yy range

90 – 200 ml/min (5Ah – C8h)

yy default

175 ml (AFh)

Example:

<F205A> set the pump flow rate to 90 ml/minute.

Note: Using this command to set a desired flow rate, the pump flow
rate will revert back to default settings upon power up. The flow rate
can be fixed in the EEPROM to a specific default setting. See
EEPROM Map for details.

<C80>

Request Software Reset

Allows the host to initiate a software reset.

Note: Duet must be in unprotect mode (See QF70).