Page 33
Host Command
Description
<Qxxy>
Function enable
Allows the host to enable or disable functions listed below.
<Qxxy>
where xx (ASCII) is the compensation and y (ASCII)
is ON/OFF status.
xx =
00h
O
2
compensation
01h
N
2
0 compensation
02h
desflurane compensation
08h
water vapor compensation
09h
BTPS compensation
0A
Baseline compensation
7F
protect mode
y =
00 binary
OFF
01 binary
ON
3Fh
report current ON/OFF status
Example:
<Q011>
compensate the CO
2
waveform data for N
2
O
<Q7F0>
turn off protected mode (to allow certain commands)
<F20yy>
Change pump flow rate
Allows the host to specify the pump flow rate in milliliters per minute.
<F20yy>
where yy (8 bit) is the current pump flow rate.
yy range
90 – 200 ml/min (5Ah – C8h)
yy default
175 ml (AFh)
Example:
<F205A> set the pump flow rate to 90 ml/minute.
Note: Using this command to set a desired flow rate, the pump flow
rate will revert back to default settings upon power up. The flow rate
can be fixed in the EEPROM to a specific default setting. See
EEPROM Map for details.
<C80>
Request Software Reset
Allows the host to initiate a software reset.
Note: Duet must be in unprotect mode (See QF70).