Host Command
Module Response
Module Response Format
<C20>
Request software version
<Vxxxmmddyyyy>
xxx = version #
mm = version month
dd = version day
yyyy = version year
Example:
<V13010231998>
version = 1.30
date = 10-23-1998
<C21>
Request hardware version
<Hxx>
xx = hardware version
Example:
<H25>
hardware version = 2.5
<C22>
Request ambient barometric
pressure reading from
absolute pressure transducer
<Lxxxx>
xxxx = ambient barometric pressure in
mmHg
Example:
<L02E9>
ambient baro pressure =745mmHg
<C23>
Request sensor temperature
<Txx>
xx = (temperature
× 4) in
o
C
Example:<TA8>
temperature = (168/4) = 42°C
<C24>
Request chopper motor speed
<Mxxxx>
xxxx = (motor period
× 10) in msec
Example:
<M012C>
chopper motor speed = 30.0 msec
<C26>
Request single CO
2
measurement
<Wxx>
xx = CO
2
in mmHg
× 256
CO
2
= 16 bit binary integer
Example:
<w26AF>
CO2 = 38.80 mmHg
<C27>
Request single breath packet
<zxxyyww>
xx = ETCO
2
in mmHg
yy = RR in bpm
ww = InsCO
2
in mmHg
Example:
<z251800>
ETCO2 = 37mmHg, RR = 24 bpm,
InsCO2 = 0
<C2E>
Request current pump flow
rate
<Bxxxx>
xxxx = pump flow rate in ml/min
Example:
<B00AF>
Flow rate = 175ml/minute
<C31>
Request sensor EEPROM
revision #
<Rxx>
xx = revision #
Example:
<R02>
revision # = 2
<C32>
Request sensor OEM specific
code
<Ixx>
xx = OEM specific code
Example:
<I00>
OEM specific code = 00
<C33>
Request sensor serial #
<Nxxxx>
xx = serial #
Example:<N34C2>
serial # = 13506