Host command/module response overview table – Welch Allyn Duet CO2 Module - User Manual User Manual

Page 44

Advertising
background image

Duet CO

2

Module OEM Implementation Manual

Page 44

Confidential

Welch Allyn OEM Technologies

Host Command/Module Response Overview Table

Command

Command
Description

Response

Response
Description

<M04>

Enter MS auto run mode

<Sxx06>

Acknowledge with module status

<M24>

Enter SS auto run mode

<Sxx06>

Acknowledge with module status

<M03>

Enter MS measurement mode

<Sxx06>

Acknowledge with module status

<M23>

Enter MS measurement mode

<Sxx06>

Acknowledge with module status

<M02>

Enter MS warm standby mode

<Sxx06>

Acknowledge with module status

<M01>

Enter MS standby mode

<Sxx06>

Acknowledge with module status

<M21>

Enter SS standby mode

<Sxx06>

Acknowledge with module status

Invoke with enter measurement or
auto run mode command

<Wxxyy>

CO

2

waveform data packets sent

at programmed interval

Invoke with enter measurement or
auto run mode command

<Zxxyyww>

Breath data packets sent at
programmed interval

<C00>

Request module status

<Sxxyy>

Acknowledge with module status

<C20>

Request software version

<Vxxxmmddyyyy>

Version # with month, day and
year sent

<C21>

Request hardware version

<Hxx>

Hardware version # sent

<C22>

Request ambient barometric
pressure

<Lxxxx>

Ambient barometric pressure sent

<C23>

Request sensor temperature

<Txx>

Sensor temperature sent

<C24>

Request chopper motor speed

<Mxxxx>

Chopper motor speed sent

<C26>

Request single CO

2

measurement

<Wxx>

CO

2

measurement sent

<C27>

Request single breath data packet

<zxxyyww>

Breath data packet sent

<C2E>

Request current pump flow rate

<Bxxxx>

Current pump flow rate sent

<C31>

Request sensor EEPROM
revision #

<Rxx>

Revision # sent

<C32>

Request sensor OEM specific

<Ixx>

Sensor OEM specific code sent

<C33>

Request sensor serial #

<Nxxxx>

Sensor serial # sent

<C3A>

Request last calibration date

<Dmmddyyyy>

Last calibration date sent

<C80>

Request software reset

No response

<C82>

Clear sensor error

No response

<Axx>

Specify breath data timeout

<axx>

Echo data field

<F20yy>

Specify pump flow rate

<f 20yy>

Echo command

<Nxx>

Specify CO

2

waveform data

packet interval

<nxx>

Echo data field

<Oxx>

Specify breath data packet interval

<oxx>

Echo data field

<Qxxy>

Specify measurement
compensation ON or OFF, SW
reset

<qxxy>

Echo new and previous
compensation settings

Advertising