Low Dead Space Single-Use Airway Adapter Specifications
PHYSICAL SPECIFICATIONS
Overall Dimensions
2.04"L.x 0.88"W x 1.14"H
(5.2cm L x 2.24cm W x 2.9cm H)
Contributory Dead Space
<1cc (1ml)
Fittings:
Male O.D. 15mm, per ISO 5356-1 and ASTM F1054
Female I.D. 15mm, per ISO 5356-1 and ASTM F1054
Resistance:
See Figure B
Respiration Rate
0 to 150 bpm
Nominal Weight:
4.8g
Material:
Polycarbonate resin and sapphire
Pressure Leak Tested at:
15psi
Environmental Conditions
Ambient Shipping/Storage:
Temperature
Relative Humidity
Barometric Pressure
-4° to 140° F (-20° to 60° C)
15-95% RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)
Ambient Operating Conditions:
Temperature
Relative Humidity
Barometric Pressure
50° to 140° F (10° to 60° C)
15%-95%RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)
Note:
This airway adapter is designed for a maximum flow of 30 l/min. This airway
adapter is not designed for reuse on another patient.
Pressure Drop
(cm H
2
0)
0
2
4
6
8
10
0
5
10
15
20
25
30
Flow (Liters/min)
Figure B
Average Pressure Drop Across the Low Dead Space Airway Adapter