Duet CO
2
Module Interface
Power Requirements
Description
Typical
Input voltage range
+8.00 to +12.00 VDC, 150 mV peak to peak ripple
a
Non-measurement power
570 mW at +8.00 VDC
Mainstream warm up power
5 W at +8.00 VDC
Mainstream operating power
2.8 W at 25° C, average
Sidestream operating power
1.3 W at +8.00 VDC
Sidestream occluded power
1.8 W at +8.00 VDC
Sidestream pump stalled power
3.0 W at +8.0 VDC
a. Optimal input voltage is +8.00 VDC. Typical power consumption increases slightly with higher input
voltages.
The host must supply fused input voltage to the Duet CO
2
Module. The actual slo blo fuse values for
the power input side of the Duet CO
2
Module are 1.00 A (12.00 V input voltage), 1.25 A (10.00 V input
voltage) and 1.50 A (8.00 V input voltage).
CO
2
Main Processor Board
Description
Specification
Dimensions
(MS & SS included)
3.15 in L
× 3.90 in W × 1.22 in H
(80 mm L
× 100 mm W × 31 mm H)
Clearance
1.00 in (25.40 mm) above board
0.20 in (5.00 mm) below board
Orientation
No limitations
Weight
< 1 lb (454 gm)