Module OEM Implementation Manual
Page 36
Confidential
Welch Allyn OEM Technologies
Mode Command Responses
Host Command
Module Response
Module Response Format
Enter Autorun Mode
<M04> for Mainstream
<M24> for Sidestream
<Sxx06>
xx = mode
yy = message
Example:
<SC406>
C = MS and SS hardware present,
mainstream operation
4 = Autorun mode
06 = Acknowledge mode command
Enter Standby Mode
<M01> for Mainstream
<M21> for Sidestream
<Sxx06>
xx = mode
yy = message
Example:
<SE106>
E = MS and SS hardware present,
sidestream operation
1 = Standby mode
06 = Acknowledge mode command
Enter Warm Standby Mode
<M02> for Mainstream (only)
<Sxx06>
Example:
<SC206>
C = MS and SS hardware present,
mainstream operation
2 = Warm Standby mode
06 = Acknowledge mode command
Enter Measurement Mode
<M03> for Mainstream
<M23> for Sidestream
<Sxx06>
Example:
<SE306>
E = MS and SS hardware present,
sidestream operation
3 = Measurement mode
06 = Acknowledge mode command
Simple Command Response
Host Command
Module Response
Module Response Format
<C00>
Request module status
<Sxxyy>
xx = mode
yy = message
Example:
<S6400>
6 = MS hardware out, SS hardware
present, sidestream operation
4 = Autorun mode
00 = OK status