Welch Allyn Duet CO2 Module - User Manual User Manual

Page 36

Advertising
background image

Duet CO

2

Module OEM Implementation Manual

Page 36

Confidential

Welch Allyn OEM Technologies

Mode Command Responses

Host Command

Module Response

Module Response Format

Enter Autorun Mode
<M04> for Mainstream
<M24> for Sidestream

<Sxx06>

xx = mode
yy = message

Example:
<SC406>
C = MS and SS hardware present,
mainstream operation
4 = Autorun mode
06 = Acknowledge mode command

Enter Standby Mode
<M01> for Mainstream
<M21> for Sidestream

<Sxx06>

xx = mode
yy = message

Example:
<SE106>
E = MS and SS hardware present,
sidestream operation
1 = Standby mode
06 = Acknowledge mode command

Enter Warm Standby Mode
<M02> for Mainstream (only)

<Sxx06>

Example:
<SC206>
C = MS and SS hardware present,
mainstream operation
2 = Warm Standby mode
06 = Acknowledge mode command

Enter Measurement Mode
<M03> for Mainstream
<M23> for Sidestream

<Sxx06>

Example:
<SE306>
E = MS and SS hardware present,
sidestream operation
3 = Measurement mode
06 = Acknowledge mode command

Simple Command Response

Host Command

Module Response

Module Response Format

<C00>
Request module status

<Sxxyy>

xx = mode
yy = message
Example:
<S6400>
6 = MS hardware out, SS hardware
present, sidestream operation
4 = Autorun mode
00 = OK status