CAUTIONS AND WARNINGS


To prevent injury and damage to your scale, please follow these instructions very carefully.

Do not transport the scale while the platform is loaded. Unplug the adapter and store both the

adapter and cord on the adapter holder before moving the scale. Be sure the scale is tilted

enough to assure it is rolling properly on its wheels when it is moved for cleaning or transported

from one room to another.

For accurate weighing the scale must be placed on a flat, stable surface.

For accurate weighing, verify proper operation according to the procedure described in this manual

before each use.

Do not use in the presence of flammable materials.

Operating at other voltages and frequencies than specified could damage the equipment.

If the “LOW BAT” indicator activates, for accurate weighing, replace the batteries or connect the

scale to a power source, using the power adapter, as soon as possible.

SPECIFICATIONS

GENERAL

The ProPlus

®

2101KGWA Digital Handrail Scale uses highly sophisticated microprocessor

technology. Each precision scale is designed to provide accurate, reliable and repeatable weight

measurements and features that make the weighing process simple, fast and convenient.

This scale is set up to use motion-sensing weighing technology, to determine the actual weight of a

moving patient. The scale may be changed to measure live weight (see the ProPlus

®

Display

Module User Manual for instructions on how to change the scale settings).

Weight is displayed in kilograms. The scale features a wrap-around handrail for patient comfort

and safety. It also has a 180° swivel display head for privacy and easy operator reading from

either side of the scale. It has two wheels built in for easy mobility. The scale can be operated

using its power adapter (included) or by 6-D cell batteries (not included), depending on use—see

Connectivity section, below.

2101KGWA SPECIFICATIONS

0B

Capacity and
Resolution

2101KGWA: Max 400kg / Min 1kg, e=0.1kg (Class III)

1B

Power Requirements 120V-240V AC, 50-60Hz - 9V DC adapter, Adapter Model ADPT30

(INCLUDED) or 6 D-cell batteries (not included)

2B

Environmental

Operating temperatures: 10°C to 35°C (50°F to 95°F)
Storage temperatures: 0°C to 50°C (30°F to 125°F)
Maximum Humidity: 85% RH

3B

Physical Dimensions Platform

Length: 56 cm (22”)
Width: 66 cm (26”)
Height: 5.7 cm (2-¼”)

Assembled Product Size and Weight

Length: 76 cm (30”)
Width: 70.5 cm (27-¾”)
Height: 125 cm (49-2/5”)
Weight: 32.5 kg (72 lb)