Nextteq VeriFit Irritant Smoke Generators for Respirator Fit Testing User Manual

Page 23

Advertising
background image

Page 20 of 24

VeriFit Irritant Smoke Generators Manual

Toll free: 877-312-2333 / Toll free: 877-312-2444 www.nextteq.com

PN 50811300-340N Rev F 5/09

Operating Instructions

11. A test subject’s sensitivity to irritant smoke varies based upon size,

weight, heredity, age, health, and other factors. Some test subjects

may respond to a greater degree than others to irritant smoke.

During the Sensitivity Check, the administrator shall consider

the test subject’s sensitivity and be careful to generate a small,

dilute, weak stream of smoke that is the absolute minimum

concentration of irritant smoke necessary to elicit a response from

the test subject.

12. Do not re-expose the test subject to irritant smoke until the test

subject specifically and expressly confirms that all effects from

the prior exposure to irritant smoke have ceased.

13. During the Sensitivity Check, do not place the open end of the

smoke generator closer than 36 inches from the test subject’s face.

14. This product(s) is designed for qualitative respiratory fit testing

in ambient atmospheres of 40% relative humidity to 90% relative

humidity. Do not use this product(s) in atmospheres of less than

40% relative humidity.

15. Test only one test subject at a time.
C. Smoke Generation and Flow Rate
1. Ensure you are in compliance with all of the previous steps,

which are required prior to smoke generation.

2. The VeriFit Irritant Smoke Generator has an internal volume of

approximately 8 milliliters. The pump produces a volume of

approximately 7 milliliters with each complete and uniform

bulb compression.

3. To produce a flow rate of approximately 200 milliliters per minute

with the VeriFit Irritant Smoke Generator, uniformly and

completely compress the bulb smoothly and continuously every

2 seconds for 1 minute.

4. To produce a small, weak dilute concentration of smoke

using the VeriFit Irritant Smoke Generator, slowly and gently

compress the bulb at a controlled, uniform, steady rate until a

small, weak, diluted stream of smoke is produced that is the

absolute minimum necessary to elicit a response from the test

subject being tested. Then immediately stop compressing the

bulb and release the partially compressed bulb. No additional

smoke is to be generated.

Advertising