Pressure monitor (psi), Temperature monitor (°c), Operator’s manual – ANKOM TDF Dietary User Manual

Page 69

Advertising
background image

Operator’s Manual


Rev E 12/2/14

pg. 69

The values shown on this screen will dynamically change as the temperatures change. When the

Display Pressures button

is pressed, the following screen is displayed.

The readings shown on this screen for the input pressure and the lines to all six stations represent the absolute pressure.
These values will dynamically change as the pressures change.

14. Start new procedures

Button #14 is displayed when a top process completes. At that time, the button reads: “Start New Procedure?” Button #14
is also displayed when a bottom process completes and no top process is in progress. At that time, the button reads: “Main
Menu.” When button #14 is pressed, the appropriate screen for starting a new procedure is displayed.

15. Filter time increment

Button #15 is displayed when a top or bottom process begins its filter action. At that time, the button reads: “Filter Time
+.” When button #15 is pressed, the remaining filter time is increased by one minute for every time the button is pressed
up to a limit of 54 additional minutes.

16. Filter bypass or Precipitation bypass

Button #16 is displayed when a top or bottom process begins its filter action or when a bottom process begins its
precipitation phase. During the filter actions, the button reads: “Filter bypass.” During a precipitation phase, the button
reads: “Precipitation bypass.” When pressed, button #16 instructs the instrument to stop the filter or precipitation action
and start the next action.

If you mistakenly increase your filter time using the Filter time increment button,
you can use the Filter bypass to stop the filter process before the additional filter
time occurs.

17. Vacuum sensor pressure

The value shown in box #17 is the current pressure (in psi units) read by the vacuum sensor. If an error occurs during
operation, and the user overrides the error to continue the operation, this box will not display a value.

Pressure Monitor (psi)

xx.x

Line 1

xx.x

Line 2

xx.x

Line 3

xx.x

Line 4

xx.x

Line 5

xx.x

Line 6

xx.x

Input Pressure

xx.x

Vacuum Sensor

Temperature Monitor (°C)

XX

Paddle 1

XX

Paddle 2

XX

Paddle 3

XX

Paddle 4

XX

Paddle 5

XX

Paddle 6

XX

In-line Heater