QC / Calibration Checks

The ANKOM

TDF

Dietary Fiber Analyzer is an automated and highly integrated instrument. Utilizing an on-board computer, the

instrument automatically checks its operation in multiple ways. For example, during each run the instrument automatically
checks for Tube Integrity.

In addition to reviewing assay results, you can determine if the instrument is functioning properly by executing the following
procedures.

Fluid Delivery

You will use the Volume Calibration diagnostic to check and calibrate the Fluid Delivery system. This diagnostic allows you to
verify that the instrument is delivering the correct amount of fluid through the pump. This calibration should be done after
installing new pump tubing. To run the Volume Calibration procedure, follow the steps below.

1. Press the

Diagnostics button on the screen above. The following screen will be displayed.

2. Press the

Service Mode button. The following screen will be displayed.

Diagnostics

Touch Screen Display version: A.AA
SPLat controller software version: vX.XX dd-mm-yy

Pump Tube

Test

Expert Mode

(OFF)

Faults

Service

Mode

Clear IDF /

SDF lines

Digestion

Times

Heater Test

Temperatures

Pressures

Line Charge

Motor Test

Valve Test

Select a Function

AOAC 991.43

IDF/SDF

AOAC 985.29

TDF

Diagnostics

AOAC 991.43

TDF