Volume calibration, Start, Service mode – ANKOM TDF Dietary User Manual

Page 84

Advertising
background image

Operator’s Manual


pg. 84

Rev E 12/2/14

3. Press the

Volume Calibration button. The following screen will be displayed.

4. Label and weigh six small empty glass beakers.

5. Press the

START button on the Volume Calibration screen above. The following screen will be displayed.

Place spare containers under IDF Delivery
Nozzles to catch pre-calibration charge.

Press OK when done.

OK

Volume Calibration

Volume Calibration

xx.x

Calibration Time

START

Service Mode

Temperature

Sensor

Corrections

IDF Count

Volume

Calibration

Bypass Amylase

Phase Initial

Heat-up

(NO)

AOAC 2009.01/

2011.25

Registration