Pump tube test start, Operator’s manual – ANKOM TDF Dietary User Manual

Page 76

Advertising
background image

Operator’s Manual


pg. 76

Rev E 12/2/14

Digestion Times

When you press the

Digestion Times button the following screen will be displayed.

This diagnostic allows you to modify the times used by the computer for the digestion processes. This can be helpful when
troubleshooting or demonstrating the instrument. If you press the

Save to Permanent Memory button, the values you entered

will be saved until you change them again by using this diagnostic. If you do not press the

Save to Permanent Memory button,

the values you entered will be saved until power is turned off, or until you change them again by using this diagnostic.

If you start a new procedure with Digestion Times that are less than the official
method times, a message will be displayed on the Touch Screen Display that allows
you to reset the values.

Pump Tube Test

When you press the

Pump Tube Test button on the Diagnostics screen, the following screen will be displayed.

This diagnostic tests the integrity of the tubing from the pump to the inlet valves (IDF, SDF, and Waste). When you press the
START button, the following screen will be displayed.

All lines should be charged before test is run.
Verify that all supply containers are connected and
contain the desired fluid.

Cancel

Continue

(no line charge)

Continue

(line charge)

Pump Tube Test

START

Status

Pressure

difference

(psi)

1

2

3

4

5

6

AOAC 991.43/985.29 Digestion Times

(Press button with number to change)

60

Amylase

Digestion Time

(minutes)

SDF Precipitation Time

(minutes)

20

Protease

Digestion Time

(minutes)

30

HCl/AMG

Digestion Time

(minutes)

30

Reset to official

method times

Save to

Permanent

Memory