Putting into operation 5.1, Installation site, Electrical connections – BUCHI KjelFlex K-360 User Manual

Page 23: 5putting into operation

Advertising
background image

5 Putting into operation

23

KjelFlex K-360 Operation Manual, Version C

5

Putting into operation

This chapter describes how the instrument is installed and gives instructions on initial startup.

NOTE
Inspect the instrument for damages during unpacking. If necessary, prepare a status report immedia-
tely to inform the postal company, railway company or transportation company.

Keep the original packaging for future transportation.

5 .1

Installation site

Install the instrument on a clean, flat and stable base and place the tanks for chemicals beside the
instrument (not higher and not more than 1 m lower).

Notice

Reduced instrument safety and lifetime due to improper handling and installation location.

Do not place any object on top of the instrument.

Keep a free safety space of 30 cm around the instrument.

Do not place items behind the instrument.

Do not operate the instrument within an fume cabinet.

5 .2

Electrical connections

L

N

IEC-60320, Type: C13

Fig. 5.1: Electrical connections

!

WARNING

Death or serious burnings by electric shock.

Mains voltage must match the voltage reading on the type plate.

Instrument must be earthed by the mains socket.

Only use a molded 3-pole device plug (IEC-60320, type C13 or C15).

Do not use damaged cables.

Advertising