Troubleshooting 8.1, Malfunctions and their remedy, 8troubleshooting – BUCHI KjelFlex K-360 User Manual

Page 54: 8 .1 malfunctions and their remedy

Advertising
background image

8 Troubleshooting

54

KjelFlex K-360 Operation Manual, Version C

8

Troubleshooting

This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a Buchi technical
engineer who has access to the official service manuals. In this case, please refer to your local Buchi
customer service agent.

8 .1

Malfunctions and their remedy

Table 8-1: General malfunctions and their remedy

Error number / Error indication

Possible cause

Corrective measure

0

The current process was stopped.

Press OK, aspirate manually and start
again

1

Steam Valve (Y1 on) malfunction.

Contact the Buchi customer service

2

Aspiration Out Valve (Y2 on) malfunction.

Contact the Buchi customer service

3

Aspiration In Valve (Y3 on) malfunction.

Contact the Buchi customer service

4

Receiver Valve (Y4 on) malfunction.

Contact the Buchi customer service

5

Cooling Water Valve (Y5 on) malfunction.

Contact the Buchi customer service

8

Injection Valve (Y8 on) malfunction.

Contact the Buchi customer service

9

Water Valve (Y9 on) malfunction.

Contact the Buchi customer service

11

Steam Valve (Y1 off) malfunction.

Contact the Buchi customer service

12

Aspiration Out Valve (Y2 off) malfunction.

Contact the Buchi customer service

13

Aspiration In Valve (Y3 off) malfunction.

Contact the Buchi customer service

14

Receiver Valve (Y4 off) malfunction.

Contact the Buchi customer service

15

Cooling Water Valve (Y5 off) malfunction.

Contact the Buchi customer service

18

Injection Valve (Y8 off) malfunction.

Contact the Buchi customer service

19

Water Valve (Y9 off) malfunction.

Contact the Buchi customer service

20

H

2

O Pump (M1 on) malfunction.

Contact the Buchi customer service

21

NaOH Pump (M2 on) malfunction.

Contact the Buchi customer service

22

Reagent Pump (M3 on) malfunction.

Contact the Buchi customer service

23

H

3

BO

3

Pump (M4 on) malfunction.

Contact the Buchi customer service

25

H

2

O Pump (M1 off) malfunction.

Contact the Buchi customer service

26

NaOH Pump (M2 off) malfunction.

Contact the Buchi customer service

27

Reagent Pump (M3 off) malfunction.

Contact the Buchi customer service

28

H

3

BO

3

Pump (M4 off) malfunction.

Contact the Buchi customer service

30

Calibration out of range, check pump.

Check connection K-360/tank, check pump

33

The service door is open. If a process was
in progress, it may have been stopped.

Close the service door, start again

34

The splash protector door is open. If a
process was in progress, it may have been
stopped.

Close the splash protector door, start again

Advertising