9 Shutdown, storage, transport and disposal

58

KjelFlex K-360 Operation Manual, Version C

9

Shutdown, storage, transport and disposal

This chapter instructs how to shut down the instrument, how to pack it for storage or transport, and
specifies the storage and shipping conditions.

9 .1

Preparing the instrument for transport

Take out the power cord and remove all water/reagent hoses. Rinse the sodium hydroxide pump/
reagent pump/boric acid pump hose with distilled water. Clean the instrument thoroughly, so that all
chemical residues are removed completely.

!

WARNING

Death or serious injuries by electric shock.

Switch off the instrument and remove the power cord.

•

Do not bring any electrical component into contact with fluids.

•

Do not operate the instrument when enclosure is open or removed.

•

!

CAUTION

Risk of burnings by hot machine parts and water steam.

Let the system and the steam generator cool down for 30 minutes before performing any

•

service.
Do not operate the instrument when enclosure is open or removed.

•

!

WARNING

Death or serious burnings by flammable vapors.

Remove all sources of flammable vapor.

•

Avoid spark formation.

•

Do not operate with open flames.

•

!

WARNING

Serious chemical burnings by corrosives.

Observe supplementary data sheets of all used chemicals.

•

Wear protective goggles.

•

Wear protective gloves.

•

Wear protective clothing.

•

To empty the steam generator, proceed as follows:

Turn off the instrument.

•

Let the steam generator cool down for 30 minutes.

•