7 Maintenance

50

KjelFlex K-360 Operation Manual, Version C

NOTICE

Risk of sealing damage by sharp edges or lubricants.

When installing, move sealing in perpendicular axis to the glass parts.

•

Do not touch sealing with potentially sharp objects.

•

Do not apply grease or any other lubricant to the sealing.

•

For further information on how to unmount and remount the splash protector, see chapter 7.3.3.

7 .3 .2

Replacing the splash protector

Replace the Splash protector after approximately 5000 determinations.

7 .3 .3

Replacing the rubber bung seals and the splash protector

We recommend to replace the parts listed below depending on the instrument use, at least once a
year.

The seals are subject to wear and tear, thus you should check them regularly. We recommend to
replace them once a year. The following parts need to be replaced (see also the spare parts list):

1

2

3

a Accessories splash protector
b Gasket for splash protector, complete
c SVL 22 sealing

Make sure that the instrument is cooled down.
Open the service door and loosen the two tubings
shown on the picture.