6 Operation

36

KjelFlex K-360 Operation Manual, Version C

6 .4 .1

Pump calibration

Go to Pump Calibration and press Enter. The following screen appears:

Fig. 6.3: Pump calibration

It is recommended to calibrate the pumps with the same volume as used for the methods. To perform
the calibration a graduated cylinder is needed.
To carry out the pump calibration, proceed as follows:

Choose the required pump by pressing left / right arrow repeatedly, when e.g. “H

2

0” is highlighted.

Enter the quantity to be dosed (Dose Volume), e.g. 50 ml.

Press Run to start dosing.

Measure the dosed volume and enter the measured volume.

Press Cal to calibrate the pump. When the calibration is finished a corresponding calibration factor

appears.

NOTE
Repeat this procedure until the entered and the dosed volume correspond. An acceptable difference
at 50 ml is ±2 ml. NaOH/H

2

O/Reagent can be dosed into the sample tube and then poured into a

graduated cylinder. H

3

BO

3

is dosed directly into the receiving vessel and then poured into a gradu-

ated cylinder.

It is recommended to repeat the pump calibration regularly (once a month or every 100 distillations).

6 .4 .2

Setting date and time

Go to Set Date and Time and press Enter. The following screen appears:

Fig. 6.4: Date/Time

Enter the corresponding settings via keypad and press Save to store them.