6 Operation

36

KjelFlex K-360 Operation Manual, Version C

6 .4 .1

Pump calibration

Go to Pump Calibration and press Enter. The following screen appears:

Fig. 6.3: Pump calibration

It is recommended to calibrate the pumps with the same volume as used for the methods. To perform
the calibration a graduated cylinder is needed.
To carry out the pump calibration, proceed as follows:

Choose the required pump by pressing left / right arrow repeatedly, when e.g. “H

•

2

0” is highlighted.

Enter the quantity to be dosed (Dose Volume), e.g. 50 ml.

•

Press Run to start dosing.

•

Measure the dosed volume and enter the measured volume.

•

Press Cal to calibrate the pump. When the calibration is finished a corresponding calibration factor

•

appears.

NOTE
Repeat this procedure until the entered and the dosed volume correspond. An acceptable difference
at 50 ml is ±2 ml. NaOH/H

2

O/Reagent can be dosed into the sample tube and then poured into a

graduated cylinder. H

3

BO

3

is dosed directly into the receiving vessel and then poured into a gradu-

ated cylinder.

It is recommended to repeat the pump calibration regularly (once a month or every 100 distillations).