Message of status and error – Checkline FGS-220VC User Manual

Page 29

Advertising
background image

- 24 -

6. Message of status and error




The lower limit was reached.
“PUSH LMT” is blinking.

If the status is abnormal, move the
lower limit switch position.




The upper limit was reached.
“PULL LMT” is blinking.

If the status is abnormal, move the
upper limit switch position.




The over load of the mounted force gauge
occurred in push direction.

Confirm whether the force gauge
used is correct.
Check whether the measurement
is abnormal.




The over load of the mounted force gauge
occurred in pull direction.

Confirm whether the force gauge
used is correct.
Check whether the measurement
is abnormal.




The emergency stop was pushed.

Confirm that a problem is not
found, then release the emergency
stop.




The motor or the driver of the motor was
abnormal.

Turn off and wait one minute, then
turn on.
Contact TECHNICAL SUPPORT if
the alarm is not canceled.




The communication error between the
controller and the motor driver in the stand
occurred.

Turn off and wait one minute, then
turn on.
Contact to our technical support if

the error alarm is not canceled.






The read error of the memory in the stand
occurred.

Turn off and wait one minute, then
turn on.
If the error is not canceled,
initialize the parameter. Refer to
the below procedure of the
initialization.






The write error of the memory in the stand
occurred.

Turn off and wait one minute, then
turn on.
If the error is not canceled,
initialize the parameter. Refer to
the below procedure of the
initialization.






The stand could not recognize the
mounted force gauge.

Check:
The parameters setting of the force
gauge, the sequence of power on
and the cable.
If no problem, ask to our technical
support.

22

Advertising