2 dry sieving, 1 sieving parameters, 2 sieving aids – FRITSCH PULVERISETTE 0 User Manual

Page 29: Accessories

Advertising
background image

Maximum load:
‐ Sieving stock < 1 kg
‐ Sieves and sieving stock < 6 kg

7.2.2 Dry sieving

For dry sieving, use the dry sieve pan (without outlet) and the dry

clamping lid (plexiglas clamping lid without nozzles).

7.2.2.1

Sieving parameters

Parameters

Coarse sample

Fine sample

Sieving time

3...20 min

15....30 min

Amplitude

2.5....3 mm

1.5....2.5 mm

In order to achieve an amplitude set point of 3 mm (full deflection), at

least 3 sieves, collecting vessels and clamping lids must be clamped on

the sieve shaker. If fewer sieves are clamped, it may not be possible to

achieve an amplitude set point of 3 mm.
In this case, the sieve shaker controller can no longer set the optimal

working point; the sieve shaker vibrates with a lower amplitude and fre-

quency. The amplitude set point must be reduced accordingly in this

case. When the machine has warmed up, the set point can be increased

accordingly.

7.2.2.2

Sieving aids

To shorten the sieving time, sieving aids can be attached in each of the

sieves that have mesh widths larger than 32 µm. During sieving, the balls

jump on the sieves and accelerate the output of the sieving stock.
The following sieving aids:

n

Agate balls
5 mm Ø: 15 pieces per sieve or

n

Agate balls
10 mm Ø: 10 pieces per sieve or

n

Rubber balls
20 mm Ø: 5 pieces per sieve or

can be used.