9 maintenance, Maintenance, 9maintenance – FRITSCH PULVERISETTE 13 premium line User Manual

Page 38

Advertising
background image

9

Maintenance

DANGER!
Mains voltage
– Before beginning with maintenance work, unplug the

mains plug and protect the device against being unin-

tentionally switched back on again!

– Indicate maintenance work with warning signs.
– Maintenance work may only be performed by special-

ised personnel.

– Put safety equipment back into operation after mainte-

nance or repair work

We recommend keeping a safety logbook Ä Chapter 15

‘Safety logbook’ on page 54, where all work (maintenance,

repairs......) performed on the device is entered.

The most important element of maintenance is regular

cleaning:

When cleaning the complete device, adhere to the

guidelines of the Accident Prevention Regulation (BGV

A3) ‐ especially if the device has been set up in a dusty

environment or if the processed source material pro‐

duces dust.

Functional part

Task

Test

Maintenance interval

Grinding chamber sensor

Start-up block

Open the grinding

chamber; start; error mes-

sage

Before each use

Grinding chamber safety

lock

Start-up block

Do not close grinding

chamber by motor; start;

error message

Before each use

Grinding stock container

protective equipment

Start-up block

Grinding stock container

not inserted; start; error

message

Before each use

Grinding disks gap width

Break function

Test distance

Before each use

Maintenance

- 38 -