16.

Open the grinding chamber and slide the locking spanner as far as

possible through the hinge of the grinding chamber.

Depending on the wear on the old grinding disks, the

position of the retracted movable grinding disk may

mean that the locking spanner will not fit behind it

any more. From point 14, positioning behind the mov‐

able grinding disk is no longer necessary. The locking

spanner should be used to secure the door of the

grinding chamber for simpler handling when inserting

the fixed grinding disk and the funnel.

17.

Insert the fixed replacement grinding disk and fixate it with the

hexagon screws. Fasten the disk with the locking handle (K).

Be careful when opening it to insert the fixed replace‐

ment grinding disk.

18.

Insert the funnel again in the reverse order of what is shown

under point 11 and screw it tight with 3 cylinder screws. (See point

9 to 11)

19.

Remove the locking spanner (B) again and close the grinding

chamber manually.

20.

Connect the mains plug, turn the device on and lock the grinding

chamber by pressing the locking button (h).

21.

The zero point needs to be specified again. (See Ä Chapter 6.3

‘Setting the zero point’ on page 25).

Material

Reference

value for max.

tightening

torque (Nm)

Density g/cm³ Abrasion

resistance

Use for mate-

rial to be

crushed

Order number

fixed disk:

Order number

movable disk:

Tempered

steel cast

11-12% Cr

at least 50

7,9

Good

Hard, brittle

sample mate-

rial

13.1100.09

13.1110.09

Manganese

steel 12-13%

Mn

at least 50

7,9 - 8

Good

Hard, brittle

sample mate-

rial

13.1120.23

13.1130.23