Counting mode – A&D Weighing SJ-30KWP User Manual

Page 12

Advertising
background image

10

7. COUNTING MODE

Determines a unit weight (the weight of one piece) from a known sample quantity, and
calculates how many pieces are on the weighing pan using the unit weight.
The unit weight is maintained even if the power is turned OFF.


1. Tap the

UNITS

switch to se lect “pcs”.

(“pcs” = pieces)

2. Tap the

SAMPLE

switch to enter the sample

unit weight storing mode.
The numerical value on the left indicates the
number of samples.

3. Tapping the

SAMPLE

switch allows you to

change the number of samples in the order 5
10 20 50 100 ESC 5.
By tapping the

UNITS

switch when “

e5C

” is

displayed, the scale exits the sample unit
weight storing mode to the count display.

4. When “

-

” appears at the right side of the number of

samples, tap the

switch to zero the

scale.
If necessary, place a container on the weighing
pan, and tap the

TARE

switch. Confirm that

the right side of the number of samples shows
zero.

5. Place the correct number of samples on the

weighing pan or in the container.

6. Confirm that the STABLE indicator is turned on.

Tap the

UNITS

switch to calculate and store

the unit weight. Remove the samples. The
scale is set to count objects with this unit
weight.

The total weight of samples should be more than

shown below, regardless of the number of samples.

SJ-3000WP: 2.5

g

SJ-6000WP: 5

g

SJ-15KWP: 12.5

g

SJ-30KWP: 25

g

If not, the display shows “

lo ut

” and returns to

the display of step 5. Increase the number of
samples (step 3) and try again.

7. Place the objects to be counted on the weighing pan.

Confirm

zero

Number of

samples

Samples

Weighing

pan

Container

Tap

Tap

Tap

Tap

Tap

Advertising
This manual is related to the following products: