10. CALIBRATION

Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.

When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.

For regular calibration.

Note: The Legal for Trade models can not be re-calibrated if they have been

sealed.

10.1. Calibration mode

The calibration mode has the following three functions.

Gravity acceleration correction
Calibration using a weight
Restoring the factory set values

How to enter the calibration mode

Method 1:

1. Make sure that the scale is in the weighing mode.
2. Touch and hold the

SAMPLE

switch until the gravity acceleration value (in this

example, “9.7985”) appears and release the switch.

Note: The above operation is disabled for the Legal for Trade models.

Method 2:

1. Make sure that the scale is in the

weighing mode.

2. Loosen the two screws on the switch

cover and open the switch cover. The
calibration (CAL) switch is located inside.

3. Press the

CAL

switch. The scale

displays the gravity acceleration value (in
this example, “9.7985”).








Switch cover

CAL switch