Tap the

UNITS

switch

to select an item

and

Tap the

TARE

switch

to execute it.



To exit from the calibration mode, touch
the

switch or Press the

CAL

switch. The scale turns the power OFF.

10.2. Gravity acceleration correction

When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.

Note: Gravity acceleration correction is not required when the scale will be

calibrated using a calibration weight at the place where it is to be used.


1. Refer to “10.1. Calibration mode” to enter the calibration

mode. The gravity acceleration value is displayed.

2. Tap the

TARE

switch to enter the gravity

acceleration value setting mode.

3. Change the displayed value using the following switches.

UNITS

To shift the blinking digit to the right.

SAMPLE

To increase the value of the blinking

digit by one.


4. Tap the

TARE

switch. The display shows “

end

” and

returns to the newly stored gravity acceleration value.

5. When calibration using a calibration weight is to be

performed, go to step 3 of “10.3. Calibration using a
weight”.
To finish the setting procedure, touch the

switch

or the CAL switch. The scale returns to the weighing mode.