3 configuration parameters, 4 scale parameters, Configuration parameters – Flintec FT-16D V.1.21 User Manual

Page 13: Scale parameters, ] filter, 4 scale parameters [2--] scale block, ] set up

Advertising
background image

as close as possible to the independent load cell or sectional pair 1. Then press <Enter>. After some
period [321 02] appears on the display. Apply the same load to the other places in sequence. After
loading all load cells / sections the message [322 ] will be shown on the display. This means the
eccentricity error correction is performed automatically.

[322 ] Manual Adjustment

If needed the eccentricity error can get modified manually. You can change the error correction
coefficients of the DLCs in this parameter. For entering this parameter press the <Tare> key. You can
change the load cell number by the <Enter> key. For changing the coefficient of any DLC select the
number of the DLC and press the <Tare> key to view and edit the coefficient. Use <Tare> and <Zero>
keys for changing the value and then press the <Enter> key.
To exit this parameter press the <F> key and save the changes you applied by pressing <Enter>. The
<F> key is used to quit without saving the manual corrections.

[323 X] Service mode: Set Shift Coefficients manually to 1

This parameter allows you to view the weighbridge’s performance with disabled eccentricity correction for
service purposes. Using this function does not change any stored coefficient but all coefficients are
“temporarily” set to value 1.

0 = Normal operation (factory default)

1 = Service mode


After the eccentricity adjustment a scale calibration has to be done.

6.3 Configuration

Parameters

[1--] Configuration Block

[11-] Start Up

[114 X] Buzzer

0 : Disabled (factory default)

1 : The Buzzer will be actived if any key is pressed OR
any error message occurred OR any alarm is active.

[12-] Filter

In this block the proper filter values according to the operating conditions can be entered. One of the most
important features of the indicator is viewing the filter characteristics on the display and with the help of this
option, you can select the most suitable filter without leaving the setup mode.

[120 X ] Filter

For FT-16 the filter value can be set from 0 (slow) to 9 (fast), for FT-16D from 0 (slow) to 4 (fast). When
you enter this parameter and press the <Enter> key while [120 X] is shown on the display, the weight
variation can be seen on the display. The value of the filter can be changed by using the <

↑> and <↓>

keys and the weight variation for each value is shown on the display. After finalising the filter selection
you can go to the next step by pressing the <F> key.

6.4 Scale

Parameters

[2--] Scale Block

The parameters related to the weighing operation are located here.

[20-] Set Up

[200 X] Approved

0 = No

1 = Yes

[201 X] Increased Indication

0 = by pressing key

1 = Always increased indication

[202 X] Power On Zero

This parameter setting controls if after power-on the scale will get automatically zeroed if the weight is in
the percentage of the zeroing range. If the weight is not in the zeroing range the display will show [E E E
] message until you press the <Enter>.

0 = disable

1 = ± 2%

2 = ± 10%

FT-16(D) Technical Manual, Rev. 1.21 December 2008

Page 13 of 32