Appendix 4: error table, Ppendix, Rror – Flintec FT-16D V.1.21 User Manual

Page 30: Able

Advertising
background image

A

PPENDIX

4: E

RROR

T

ABLE

The FT-16(D) weight indicator has been designed as very reliable and virtually error free instrument. However if
an error occurs do not attempt to repair the equipment before you understand what caused the error. Note the
problems you have with your instrument and the error messages shown on the display. Then try to solve the
problem according to the error table given below.

Error

Code

Description

Actions to take / Possible cause

FT-16

FT-16

Under

Weight is too low

x

x

Over

Over Load

Check the load; Load cell or instrument could
be defective

x x

Adc Out

Input signal outside the input range

Check the load; Check the calibration; Load
cell or instrument could be defective

x

E E E

Load not within zero range

See parameter [202] in chapter 6.4

x

x

Err 1

x x

Err 2

ADC error respective
Load cell communication error (DLC)

Re-energize indicator; contact FLINTEC

x x

Err 3

Indicator cannot get calibrated

Check load cell and load; repeat the
calibration

x x

Err 10

EEPROM error

Configure the instrument; EEPROM defective

x

x

Err 20

Calibration error

Calibrate the indicator

x

x

Err 21

Configuration error

Configure the indicator

x

x

Err 22

Tare, CN, Total weight and the SP,
PT and ID in use error

Check SP, PT and ID entries; Check Tare,
CN and Total weight

x x

Err 27

Indicator is not calibrated

Calibrate the indicator

x

x

Err 30

Processor Error

Contact FLINTEC

x

x

Err 34

Indicator cannot get calibrated

Load cell signal is negative , very low or too
high

x x

Err 35

Calibration Error

Calibration test weight is too small; Check
load cell connections

x x

Err 37

Scale unstable

Wait until scale becomes stable; Check
ground connection

x x

Err 40

Alibi memory error

Correct parameter [801]; Alibi memory card
broken

x x

Err 41

New alibi card installed

x

x

Err 42

Alibi card serial no. is not identical

Appears when new alibi memory card is
installed (*)

x x

Err 43

Alibi card serial no. is not identical, no
print operation will be performed

Identify the alibi memory card to the indicator
(*)

x x

Err 44

Alibi card info error

x

x

Err 46

Alibi recording capacity error

x

x

Err 47

Info error from weighing board

Contact FLINTEC

x x

Err 61

U10 (eeprom) is not installed

Contact FLINTEC

x

x

E XXXX

Hardware error

Contact FLINTEC

x

x

Ed1 yy ** Communication time out

x

Ed2 yy ** Checksum error

Check DLC cables; Check DLCs and serial
number; contact FLINTEC

x

Ed3 yy ** DLC error

Check ambient temperature; replace DLC

x

Ed4 yy ** DLC is under

x

Ed5 yy ** DLC is Over

Check mechanical installation and DLC

x

Ed6 yy ** Interface board error

Replace main board or DLC board

x

Ed7 yy ** DLC addressing error

Check serial number and re-define DLC
address OR replace DLC

x

Ed0 yy ** Other DLC error

Replace DLC

x

** yy = number of DLC

FT-16(D) Technical Manual, Rev. 1.21 December 2008

Page 30 of 32